LACO PROTOCOL DESCRIPTIONS TITANTEST
122
© 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807
431 01AF Leak rate unit UINT8 R/W Leak rate unit
0 - mbarl/s
1 - Pam3/s
2 - Torrl/s
3 - sccm
4 - sccs
5 - Atm ccs
(6 - ppm)
(7 - g/a)
(8 - oz/yr)
6...8 only in Sniff
433 01B1 Anode setpoint M2 [V] UINT16 R/W Anode voltage setpoint for mass 2 (hy-
drogen) in V
434 01B2 Anode setpoint M3 [V] UINT16 R/W Anode voltage setpoint for mass 3 in V
435 01B3 Anode setpoint M4 [V] UINT16 R/W Anode voltage setpoint for mass 4 (heli-
um) in V
441 01B9 Calibration mode UINT8 R/W Calibration mode
0 = int. auto.
1 = int. man.
2 = extern
449 01C1 Valve state UINT16 R Valve state
see table / enumerations
450 01C2 Date+Time [YMDhms] UINT8[6] R/W Date and time
use only with array-index 255 (all bytes)
year (1..99), month, day,
hour (0..23), min, sec
452 01C4 Min pressure Sniff [mbar] FLOAT R/W Minimum pressure for sniff mode [mbar]
453 01C5 Max pressure Sniff [mbar] FLOAT R/W Maximum pressure for sniff mode
[mbar]
454 01C6 Lower leak rate limit UINT8 R/W Lower leak rate limit
0 = 1.0E-12 mbar*l/s
1 = 1.0E-11 mbar*l/s
2 = 1.0E-10 mbar*l/s
3 = 1.0E-9 mbar*l/s
502 01F6 Amplifier range UINT8 R Amplifier range
0 = 13 MOhm
1 = 470 MOhm
2 = 15 GOhm
3 = 500 GOhm
4 = 13 MOhm (fixed)
5 = 470 MOhm (fixed)
6 = 15 GOhm (fixed)
7 = 500 GOhm (fixed)
Command Name Data type R/W Meaning