LACO PROTOCOL DESCRIPTIONS TITANTEST™
138
© 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807
*MEASure:DIGITALIN State of the digital inputs
0x0001 Pin 6 Start / Stop
0x0002 Pin 7 Vent
0x0004 Pin 8 Zero
0x0008 Pin 9 Calibrate
0x0010 Pin 10 Para 2
0x0020 Pin 11 ---
0x0040 Pin 12 Fill complete
0x0080 Pin 13 ---
0x0100 Sniffer button
261 R
*MEASure:IFilter Filtered ion current [A] 1573 R
*MEASure:IMeas Unfiltered ion current [A] 1568 R
*MEASure:LRMAX Maximum leak rate since last inquiry via interface in selected
unit
--- R
*MEASure:MIAKP Anode-/cathode potential [V] 170 R
*MEASure:MIAP Anode potential [V] 167 R
*MEASure:MIKP Cathode potential [V] 168 R
*MEASure:MISP Suppressor potential [V] 169 R
*MEASure:OFFset Offset current [A] 1567 R
*MEASure:P1 p1 pressure (fore line) in selected unit 130 R
*MEASure:P1:ATM p1 pressure (fore line) in atm R
*MEASure:P1:MBAR p1 pressure (fore line) in mbar 131 R
*MEASure:P1:PA p1 pressure (fore line) in Pa R
*MEASure:P1:TORR p1 pressure (fore line) in Torr R
*MEASure:P2 p2 pressure (inlet) in selected unit 132 R
*MEASure:P2:ATM p2 pressure (inlet) in atm R
*MEASure:P2:MBAR p2 pressure (inlet) in mbar 133 R
*MEASure:P2:PA p1 pressure (inlet) in Pa R
*MEASure:P2:TORR p2 pressure (inlet) in Torr R
*MEASure:REC0 Outpot voltage recorder channel 1 221 R
*MEASure:REC1 Outpot voltage recorder channel 2 221 R
*MEASure:RESULTLR AutoTest result in selected unit 703 R
*MEASure:RESULTLR:ATM*cc/s AutoTest result in atm*cc/s R
*MEASure:RESULTLR:G/a AutoTest result in g/a R
*MEASure:RESULTLR:LB/yr AutoTest result lb/yr R
*MEASure:RESULTLR:MBAR*l/s AutoTest result in mbar*l/s 705 R
*MEASure:RESULTLR:OZ/yr AutoTest result oz/yr R
*MEASure:RESULTLR:PA*m3/s AutoTest result Pa*m3/s R
*MEASure:RESULTLR:PPM AutoTest result in ppm R
*MEASure:RESULTLR:TORR*l/s AutoTest result in Torr*l/s R
Command Meaning
Relates to
LD cmd. no.
Read/Set