13.4 External Calibration Sequence
1. Open test leak
2. Start calibration: *CAL
3. Wait until *STATUS:CAL? answers "WAIT_TL_STABLE"
4. Send *CAL if signal is stable
5. Wait until *STATUS:CAL? answers "WAIT_CLOSE"
6. Close test leak
7. Send *CAL
8. Wait until *STATUS:CAL? answers "WAIT_RESULT"
9. Send *CAL to accept new calibration factors
13.5 Internal Calibration Sequence
1. Start caribration: *CAL
2. Wait until *STATUS:CAL? answers "WAIT_RESULT"
3. Send *CAL to accept new calibration factors
13.6 Examples
** Feature not available in TITANTEST™ Maintenance model
Command answer
*stat? (CR) MEAS (CR) mode
*status? (CR) MEAS (CR) mode
*read? (CR) 2.876E-7 (CR) leak rate according to programmed
unit
*read:pa*m3/s? (CR) 2.876E-6 (CR) leak rate in a different unit
*start (CR) OK (CR) start measurement
*conf:trig1? (CR) 1.0E-9 (CR) retrieve trigger 1
*conf:trig1 2.0E-9 (CR) OK (CR) set trigger 1