LACO PROTOCOL DESCRIPTIONS TITANTEST™
140
© 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807
*STATus:ERRHist Error history
*STATus:ERRHist Actual entry
*STATus:ERRHist:1 Entry 1 (newest)
*STATus:ERRHist:2 Entry 2
...
*STATus:ERRHist:15 Entry 15 (oldest)
287 R
*STATus:ERRor Current number of error / warning („NO ERROR/WARNING“ if
no error / warning)
290 R
*STATus:MAINTenanceHist Maintenance history 1 to 12 2643 R
*STATus:PREAMPRESistor Currently used resistance of pre-amplifier (13M, 470M, 15G,
500G, 13M_FIXED, 470M_FIXED, 15G_FIXED,500G_FIXED)
502 R
*STATus:RANGE Measuring range (GROSS, FINE, ULTRA) status
word
R
*STATus:RESULT Status / result of auto leak test (IDLE, RUNNING, PASS, FAIL,
NONE)
702 R
*STATus:SECINMEAS Time since change of measuring mode [s] 159 R
*STATus:SERviceHist Service history 1 to 12 2641 R
*STATus:ZERO Zero (ON, OFF) 6 R
*STOp Stop (switch from Measure to Standby) 2 W
*VENt Vent inlet port 3 W
*ZERO Switch zero on 6 W
*ZERO:OFF Switch zero off 6 W
*ZERO:ON Switch zero on 6 W
Command Meaning
Relates to
LD cmd. no.
Read/Set