LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST™
13
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
Leakage rate unit mbar l/s yes yes
Pressure unit mbar yes yes
Scaling log yes yes
Decades for scaling log 4 yes yes
Display range automatic yes yes
Time axis 32 seconds yes yes
Lower display limit 1E-12 mbarl/s yes yes
Menu PIN 0 yes no
Device PIN 0 yes no
Access to CAL function On yes yes
Language English yes no
Operating mode Vacuum yes yes
Mass Mass 4 yes yes
Leakage rate fact. Factor (1) yes no
Leakage rate filter dynamic yes yes
ZERO enabled yes yes
Time after start for ZERO 10 seconds yes yes
MS-BG-Subtraction On yes yes
Alarm operating mode Trigger alarm yes yes
Alarm delay time 30 seconds yes yes
LR setpoint 1E-8 mbar l/s yes yes
Warning limit 100% yes yes
Analog output channel 1 LR mantissa yes yes
Analog output channel 2 LR exponent yes yes
Analog scaling (upper limit) -5 yes no
Analog scaling (V/decade) 1V per decade yes no
Control location All yes yes
Mode relay 1 off yes no
Mode relay 2 off yes no
Interface** RS232 yes no
** Feature not available in T
ITANTEST™ Maintenance model
Interface protocol LD yes yes
Ext. Pump Setup Evacuation Fore pump only yes yes
Ext. Pump Setup Measure mode Fore pump only
Max. pressure 0.65 mbar yes yes
Min. pressure 0.25 mbar yes yes
Gross leak protection off yes yes
Limit value, gross leak protection 1E-3 mbar l/s yes yes
Minimum volume 0 yes yes
Volume 2 yes yes
ULTRA enabled yes yes
Switching threshold ULTRA 0.5 mbar yes yes
Table 3: Factory settings
Parameters Default value Reset to default val-
ue when factory set-
tings are loaded
Included in set of
parameters