LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST
41
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
Fig. 12: Page 1 Auto Test Process Diagram**
** Feature not available in TITANTEST™ Maintenance model
Start button is
pressed"
Evacuate Test Port
Using Roughing
Timer
Stop button is
pressed
NO
In Test Mode
and Test Timer
Active?
Measure mode
Is Delay Timer
Expired
Is HE BG Limit
Reached
Extern Trigger
Enabled?
Set Output High (OK
to Fill He) (Output
#20)
Is Fill Complete
Input On/High?
(Input #12)
Start Test Timer
Record He Leak
Rate. Store Qmax
and Qlast
Is Test Timer
Expired
End Cycle and go to
Standby
To Page 2
To Page 2
NOTE: See Delay Timer
Settings Parameter
NOTE: See He BG
Setpoint Parameter
NOTE: See External
Trigger Setup Option
Yes
No
Yes
Most sensitive vacuum range reached (e.g. ULTRA)
Yes
Yes
Yes
No
Yes
No
NOTE: See Test
Timer Settings
Parameter