Table 22: Control
ID Function Unit Command
59 Set point oset [K] IN_PAR_14
67 Control in control variable X: 0 = internal / 1 = external Pt / 2 = external
analog / 3 = external serial / 5 = external Ethernet / 6 = external EtherCAT /
7 = external Pt second (only on Integral)
[–] IN_MODE_01
69 Oset source X for set point: 0 = normal / 1 = external Pt / 2 = external
analog / 3 = external serial / 5 = external Ethernet / 6 = external EtherCAT /
7 = external Pt second (only on Integral)
[–] IN_MODE_04
Table 23: Rights
ID Function Unit Command
63 Status of Base remote control keyboard: 0 = free / 1 = blocked [–] IN_MODE_00
65 Status of remote control keyboard: 0 = free / 1 = blocked [–] IN_MODE_03
Table 24: Status
ID Function Unit Command
75 Standby status: 0 = Device is switched on / 1 = Device is switched o [–] IN_MODE_02
107 Device type (Example answer: "PRO") [–] TYPE
130 Device status: 0 = OK / -1 = Fault [–] STATUS
131 Fault diagnosis; a 7-digit answer in the format XXXXXXX is output, whereby
each digit X contains fault information (0 = no fault / 1 = fault).
The following information is defined for the seven places of the answer
format:
1. Character = error
2. Character = alarm
3. Character = warning
4. Character = overtemperature
5. Character = sublevel
6. Character = higher level (alarm setting)
7. Character = external control value missing
[–] STAT
Table 25: Contact input/output
ID Function Unit Command
96 Contact input 1: 0 = open / 1 = closed
Contact module must be present
[–] IN_DI_01
98 Contact input 2: 0 = open / 1 = closed [–] IN_DI_02
100 Contact input 3: 0 = open / 1 = closed [–] IN_DI_03
102 Contact output 1: 0 = open / 1 = closed
Contact module must be present
[–] IN_DO_01
V6 PRO bath thermostats and circulation thermostats 49 / 156