ID Function Unit Command
104 Contact output 2: 0 = open / 1 = closed [–] IN_DO_02
106 Contact output 3: 0 = open / 1 = closed [–] IN_DO_03
Table 26: SW version
ID Function Unit Command
108 Control system [–] VERSION_R
109 Protection system [–] VERSION_S
110 Remote control (Command)
(remote control unit must be present)
[–] VERSION_B
111 Cooling system
(only for devices with active cooling)
[–] VERSION_T
112 Analog interface module
(interface module must be present)
[–] VERSION_A
114 RS 232/485 interface module or Profibus / Profinet
(interface module must be present)
[–] VERSION_V
116 EtherCAT interface module
(interface module must be present)
[–] VERSION_Z
117 Contact interface module
(interface module must be present)
[–] VERSION_D
118 Solenoid valve for cooling water
(solenoid valve must be present)
[–] VERSION_M_0
119 Solenoid valve for automatic filling device
(solenoid valve must be present)
[–] VERSION_M_1
120 Solenoid valve for constant level device
(Solenoid valve must be present)
[–] VERSION_M_2
121 Solenoid valve, shut o valve 1
(Solenoid valve must be present)
[–] VERSION_M_3
122 Solenoid valve, shut o valve 2
(Solenoid valve must be present)
[–] VERSION_M_4
128 External Pt100 interface 0
(module must be present)
[–] VERSION_E
Note the following instructions:
" " (blank space) can also be used for "_".
If not otherwise specified in the command, the response will always be
given in fixed point format "XXX.XX" or for negative values "-XXX.XX" or
"ERR_X". (RS 485 interface e.g. "A015_XXX.XX" or "A015_ERR_X").
The command from the control station must be made with a CR, CRLF
or LFCR.
V6PRO bath thermostats and circulation thermostats50 / 156