Connection to Circuits
20 Power Analyzer NORMA 4000, NORMA 5000
EO1111G REV G
5.5.2 With shunt
Fehler! AutoText-Eintrag nicht definiert.
Danger! Risk of electrocution! Do not touch sense terminals!
The sense terminals at the shunts are powered with the same
voltage as the power connections.
Shunts are not isolated.
Never touch the sense terminals at the shunts.
The connecting leads to the shunts should be as short as
possible in order to prevent noise voltages.
Fehler! AutoText-Eintrag nicht definiert.
Danger! Risk of electrocution!
Risk of injury when touching connections, internal circuits and
measuring devices that are not earthed.
Always adhere to the instructions regarding the sequence of
connection (see page 12).
1
HI
CURRENT
LO
10 A max
PROBE
EXT.SHUNT
10 V max
HI
VOLTAGE
LO
1000 V max
2
HI
CURRENT
LO
10 A max
PROBE
EXT.SHUNT
10 V max
HI
VOLTAGE
LO
1000 V max
3
ALL INPUTS MAX 1000V CATII TO
1
HI
CURRENT
LO
10 A max
PROBE
EXT.SHUNT
10 V max
HI
VOLTAGE
LO
1000 V max
2
HI
CURRENT
LO
10 A max
PROBE
EXT.SHUNT
10 V max
HI
VOLTAGE
LO
1000 V max
3
L3
L2
max. 10 m
Last
I
1
L1
HI LO
Guard
*1
HI LO
Guard
*1
HI LO
Guard
*1
I
2
I
3
N