CP5 • Fault messages and troubleshooting
38 CP_SEM_60-00-60.001
5.1.2 Fault messages
Display in system menu Description
possible cause/effect
Steps to be taken (please proceed in the order listed)
A fault has been detected during the self-test.
The pump fails to switch itself off (e. g. ramp-up) (in this case:
pump 1).
Cable break
Fault in circuit board Computer board (HKR 0325)
Replace cable
Replace circuit board HKR 0325
where necessary send pump to manufacturer for checking
The pump ran too fast (without being instructed to do so).
Send pump to manufacturer for checking.
A safety feature of the pump has failed.
Cable break
Hall sensor defective
Check (and where necessary replace) connector cable
from motor control to circuit board ZPR 0801
Replace motor (hall sensors)
where necessary send pump to manufacturer for
checking
The pump speed can no longer be adjusted correctly.
Shaft encoder defective
Fault in circuit board Computer board (HKR 0325)
Replace shaft encoder
Replace circuit board HKR 0325
The pump no longer allows the system to monitor it properly.
Fault in circuit board Motor control (ZPR 0801)
Replace circuit board ZPR 0801
Turn-off path faulty
Centr.Pump 1
Runaway fault
Centr.Pump 1
Runaway fault software
Centr.Pump 1
Motor monitoring/control failed
Centr.Pump 1
Shaft angle encoder fault
Centr.Pump 1
Fault in motor controller
Centr.Pump 1
Motor controller failed
Centr.Pump 1
Fault in motor controller
Centr.Pump 1