CP5 • Fault messages and troubleshooting
40 CP_SEM_60-00-60.001
Faults in the sensor modules or sensors
Display in system menu Description
➜ possible cause/effect
◗ Steps to be taken
➜ There is a module or sensor defect. (The display of the
relevant icon indicates the corresponding monitoring
function. In this case: level monitor).
◗ Replace corresponding sensor module or sensor.
➜ There is a level sensor contact failure.
◗ Check that the level sensor is correctly seated (e.g. on the
reservoir/oxygenator).
◗ Check the metal contacts of the level sensor.
◗ Replace corresponding sensor module or sensor.
➜ The signal of the bubble sensor is too strong and cannot
be processed or the sensor is not connected.
◗ Check that the bubble sensor is correctly seated.
◗ Replace corresponding sensor module or sensor.
Receive signal too strong