CP5 • Fault messages and troubleshooting
40 CP_SEM_60-00-60.001
Faults in the sensor modules or sensors
Display in system menu Description
possible cause/effect
Steps to be taken
There is a module or sensor defect. (The display of the
relevant icon indicates the corresponding monitoring
function. In this case: level monitor).
Replace corresponding sensor module or sensor.
There is a level sensor contact failure.
Check that the level sensor is correctly seated (e.g. on the
reservoir/oxygenator).
Check the metal contacts of the level sensor.
Replace corresponding sensor module or sensor.
The signal of the bubble sensor is too strong and cannot
be processed or the sensor is not connected.
Check that the bubble sensor is correctly seated.
Replace corresponding sensor module or sensor.
Module defective
Module/sensor defective
Check level sensor
Receive signal too strong