EasyManua.ls Logo

Lonza 4D-Nucleofector - Experiment Parameter Definition

Lonza 4D-Nucleofector
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12 | 4D-Nucleofector
®
Manual
Figure 2.8: Stat experiment by typing in the experiment name or click the
“load” button to load a predefined experiment (see chapter 8.4.2).
2.8.3 Experiment parameters – overview
An experiment comprises a selected set of vessels or wells
that can only be processed with defined Nucleofection
®
Parameters.
To select a single well or a set of vessels, simply touch the
wells displayed or drag along a set of vessels with your
fingetip.
After selection, there are several options for parameter
definition. Table 2.3 describes the parameters that can
be defined.
Figure 2.9: (A) Selecting single vessels by touching the screen at the
respective well position. (B/C) Multiple vessel selections by dragging across-
the screen.
Table 2.3: Entering Nucleofection
®
Conditions
Parameter Description
Sample type
(X ,Y and 96-well Unit)
1. Sample: This selection defines a normal
sample containing cells and substrate. The
selected Nucleofection
®
Program will be applied
to this position.
2. No DNA: Negative control. Nucleofection
®
Program applied to vessel with cells but without
substrate
3. No Prog.: Negative control. No Nucleofection
®
Program applied to a vessel containing cells and
substrate. In case of “Sample” or “No DNA” the
selection has only an informative purpose. The
selection of “No program” has a functional effect,
i.e. no pulse will be applied.
Cell type code Via the “Cell type code” a predefined combina-
tion of a “Pulse code” and a “Solution” can be
selected. All Lonza optimized conditions for a
cell type are available in the “Cell type code
drop down menu. Customized cell type codes
(highlighted in black) can be added in addition to
the Lonza-defined codes (highlighted in blue), for
example, after cell line optimization (for details,
see 2.13.2.3).
Solution If the recommended Nucleofector
®
Program
is not selected automatically via the “Cell type
code” option, the “Solution” can be entered
manually.
Pulse code In case the recommended Nucleofector
®
Pro-
gram is not selected automatically via the “Cell
type code” option, a “Pulse code” (e.g. FI-115) can
be entered manually. Note: For manual “Pulse
code” selection, the “Solution” has to be selected
first (see below).
Volume (LV Unit with
LV Nucleocuvette
®
Catridge)
When using the scalable cuvette version, the
LV Nucleocuvette
®
Catridge, the volume that
should be processed has to be defined. Depend-
ing on the number of reservoirs used, one or two
volume parameters must be specified. The maxi-
mum total volume that can be entered is 20 mL.
Fill mode (LV Unit with
LV Nucleocuvette
®
Catridge)
When using the scalable cuvette version, the
LV Nucleocuvette
®
Catridge, the fill mode
(selected via ”Settings”) is indicated. It can be
changed for an individual experiment on the
experiment screen.
A
B
C
Return to TOC

Table of Contents

Other manuals for Lonza 4D-Nucleofector

Related product manuals