Millenium HX RO System
3030226 Rev. D 2-8 Installation And
Initial System Start-Up
18. Review the instructions on the screen and press and hold the HEAT PROCESS
SELECT button.
WARNING:
During heat disinfection, the RO heats the internal water to approximately 176F
(80C). To avoid injury, do not touch the RO connections during the heat cycle.
19. Press the ACTIVATE FAST COOL cooling method. Default cooling method is Slow
Cool. Review the instruction on the display. Verify all instructions have been
completed. Once verified, press and hold the SINGLE HEAT button.
20. The machine starts to purge, filling the internal tank to the proper level with
product quality water.
21. Once the purge process is complete, the RO’s internal water is circulated and
heated to a target temperature of 176F (80C).