3030226 Rev. D 2-12 Installation And
Initial System Start-Up
31. Press the Feed Water Condition Alarm. Set the RO feed water quality alarm 50
TDS/100 µS above the steady state Combined Pump Feed Water Condition
noted in step 14. If the Combined Pump Feed Water Condition is greater than
or within 50 TDS/100 µS of the alarm limits of 1000 TDS/2000 µS, disable the
alarm to prevent nuisance alarms. To disable the alarm, set the TDS setting to
1001 or conductivity setting to 2001 µS. Once complete, press .
32. Press RETURN TO MAIN.
33. The RO product may be used in conjunction with hemodialysis treatment
equipment only after obtaining test results, which validate compliance with
accepted standards.
• AAMI or CSA Product Water Panel
• Microbiological Culture (colony count) results
• Endotoxin (pyrogen) level testing
34. The Millenium HX RO is now ready for use.