3030226 Rev. D 4-28 Screen Operation And Specifications
Guidance and manufacturer’s declaration - electromagnetic immunity
The Millenium HX is intended for use in the electromagnetic environment specified below. The customer
or the user of the Millenium HX should assure that it is used in such an environment.
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output lines
2 kV for power supply
lines
1 kV for input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
1 kV line(s) to
line(s)
2 kV line(s) to
earth
1 kV line(s) to line(s)
2 kV line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(>95% dip in U
T
) for
5 s
5% U
T
(>95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
(>95% dip in U
T
) for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Note U
T
is the a.c. mains voltage prior to application of the test level.