Millenium HX RO System
3030226 Rev. D 4-29 Screen Operation And Specifications
Guidance and manufacturer’s declaration - electromagnetic immunity
The Millenium HX is intended for use in the electromagnetic environment specified below. The customer or
the user of the Millenium HX should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 Mhz
3 V/m
80 Mhz to 2.5 Ghz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the
Millenium HX, including cables, than
the recommended separation distance
calculated form the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d = 1.2P
d = 1.2P 80 MHz to 800 MHz
d = 2.3P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m)
Field strengths from fixed RF
Transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
Millenium HX is used exceeds the applicable RF compliance level above, the Millenium HX should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the Millenium HX.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less then 3 V/m.