Rad-97 Chapter 4: Operation
www.masimo.com 95 Masimo
To undo an entry, press Backspace.
3. Press Enter to access the password-protected screen.
Note: The password will have to be entered every time this screen is accessed.
Option Description Factory
Default
Settings
User Configurable Settings
Power on
Profile
Sets the profile used when the
device is powered on. See
Chapter 5: Profiles on page 103.
Previous
Profile
Previous Profile, Adult,
Pediatric, Neonatal, or User
defined profile (up to 8)
All Mute
Enabled
Enables parameter Alarm Silence
menu option. See Sounds on
page 85.
Off On or Off
Lock Alarm
Volume
Sets the lowest alarm volume
level.
Off 3, 4, or Off
Screen Lock
Allows the user to lock the
touchscreen to prevent
accidental changes.
Off On or Off
USB Port
Baudrate
Sets the USB port
communication speed.
921600
9600, 19200, 38400,
57600, 115200, 230400,
or 921600
Data
Collection
Enabled
Enables or disables physical data
collection mode.
Off On or Off
Save as
Adult*
Saves current profile parameter
as the Adult Profile.
N/A
Press Save to update the
profile.
Save as
Pediatric*
Saves current profile parameter
as the Pediatric Profile.
N/A
Press Save to update the
profile.
Save as Neo*
Saves current profile parameter
as the Neonatal Profile.
N/A
Press Save to update the
profile.
Factory
Defaults
Options are restored to factory
values.
N/A Press Restore.
* See Replacing Factory Default Settings for Adult, Pediatric and Neonatal Profiles on page
107.

Table of Contents

Other manuals for Masimo Rad-97

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Rad-97 and is the answer not in the manual?

Masimo Rad-97 Specifications

General IconGeneral
Display Size7 inches
SpO2 Accuracy±2% (70-100%)
Pulse Rate Accuracy±3 bpm
Perfusion Index Range0.02% to 20%
Pleth Variability Index Range0% to 100%
SpO2 Range0% to 100%
SpHb Range0 to 25 g/dL
SpCO Range0% to 40%
Battery TypeRechargeable lithium-ion
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Humidity Range10% to 95% non-condensing
DisplayColor touchscreen
ConnectivityWi-Fi, Ethernet, Bluetooth
Power Supply100–240 VAC, 50–60 Hz
Battery LifeApproximately 4 hours
MeasurementsSpO2, Pulse Rate, Perfusion Index, Pleth Variability Index, SpHb, SpOC, SpMet, SpCO
Pulse Rate Range25 to 240 bpm
HbCO Range0-99%
MetHb Range0-99%
SpMet Range0-99%
Data StorageInternal memory
AlarmsAudible and visual alarms for all parameters