4. The display will show the selected voltage, in this case 500 V.
5. Press soft keys 2 and 3 to move the carousel left or right
through sub-modes to PI (Polarization Index).
NOTE: The full title text of the sub-mode will appear in the
secondary field for a few seconds.
6. To start the test press the TEST button.
The test will start and the clock will appear in the bottom left of
the primary screen. The test will continue for 10 minutes. The
clock will countdown the test. An interval result will appear after
1 minute.
A warning triangle will flash on screen.
NOTE: The test can be stopped at any time by pressing the TEST
button.
7. During the tests,
The primary field displays the test results.
The secondary field displays actual voltage output and the
current measured.
8. When the test has completed,
The primary field displays the test results.
Insulation Resistance Tests
www.megger.com 21