5.15.2.10 Others
Switch off gas automatically: you can decide to stop the gas flow if the unit is not
used for some time. If activated, the valve in the system closes the flow passage
after the selected time from the last measurements.
It will be automatically activated, once you restart with the measurements.
This function avoids gas consumption when the unit is not in use. Of course, this
will cause accumulation of moisture inside the instrument (catalyst,
amalgamator, cuvette, so some blanks should be necessary before to restart
with the runs.
Show markers: if activated, the points to calculate the height of the peak and the
baseline are shown in the graph of the signal
Sample numbering: If a sequence of measurements contains a number like
unique or last letter, then, it is possible to create a list of consecutive number or
keep the same number for all measurements.
- Always: if you add measurements lines from one measurement with last or
unique letter in the sample name, the sample names will have the consecutive
numbers.