Milestone Srl MA213-003 – DMA-80 EVO – User Manual
94
5.14.1 System- Parameters
Spectrometer: here the detector voltage values of the maximum light intensity
(gain) of the lamp are shown.
The first signal refers to the detector on cell 1 and 2. (dual cell configuration)
The second signal refers to detector on long cell (cell 0, more sensitive)(tri cell
configuration)
The third signal is an option and it refers to the double beam detector installed
directly in front of the lamp.
The spectrometer is calibrated every measurement short before heating of the
amalgamator. In addition, the shutter for measuring the 0% intensity is closed for
short time. After opening of the shutter, the 100% intensity is measured.
Lamp: the lamp can be turned off manually for few seconds by clicking on this
icon. This action permits to check the dark signal (offset) with 0% of light.
Cuvette furnace: here is reported the heating element of the cuvette and its
temperature.
The temperature should be stable and constant at 125°C ± 5°C to avoid
condensation inside the cells.