Milestone Srl MA213-003 – DMA-80 EVO – User Manual
101
Errors: here all software error messages are reported
Application: here all operations on measurement files are reported
Time: date and time of the action are reported
Changes: short description of the action is reported
Filter: field to fast find a key word
Usage: memory of the terminal used by trail. When it is full, you have to get it
free to continue to keep the actions saved.
Save the trail and delete it from the page
5.15.2 Settings-Settings
5.15.2.1 Analysis
Signal analysis: Set the type of software calculation to detect the quantity of
mercury present in the sample. You can choose Integral (area of the peak) or
Maximum (height of the peak)
Signal filtering: Eliminates the interferences on the peak shape
Absorbance or emission limit of cell: absorbance limit considered from each cell
of the cuvette. The peak is considered and measured only if it is within the limit
of the cell. It the peak exceed then the next peak will be considered in the next
cell.