Connect the cuvette.
Perform the catalyst initialization procedure to
prepare the catalyst tube to the analysis
See “Catalyst conditioning procedure” enclosed in
all new catalyst tubes
It is recommended to replace the catalyst and amalgamator together after around 5000 runs
6.2 Water discharge from the outlet connection
On the outlet block of the DMA unit where the mercury trap is connected it is possible to have
accumulation of water.
It is generated by the condensation of the samples especially the ones with high water content.
It is recommended to keep the block free from water in order to avoid its passage through the outlet tube
and mercury trap.
Daily or weekly, according to the use of the instrument open the PP adapter lid and with the help of a
towel, clean block.
It is recommended to replace the mercury trap after 1 year from the installation