Milestone Srl MA213-003 – DMA-80 EVO – User Manual
121
5.16.1 Service-Autosampler
In this page the Administrator can check and in case calibrate the autosampler plate if the position are not
aligned well with the pneumatic arms.
1) Connect the gas carrier to the DMA-80 EVO unit and place the autosampler plate on the drive
disc verifying that the positions 1 market on the drive disc and autosampler plate fit together.
2) In the software, open the main menu from the top left corner of the screen and go to
Settings/Service/Autosampler
3) Press on “Cylinder II and then Cylinder I icons to move the horizontal arm completely out from the
catalyst oven.