4) Then, press “Cali” button: the autosampler plate starts to rotate and it stops when the hypothetic
position 1 of the plate is located above the vertical cylinder III.
At the same way, the software indicates the correspondent position of the autosampler plate directly in
front of the operator.
If the autosampler motor is properly calibrated, then the position 1 of the plate is in front of the catalyst
tube entrance and the position 11 is in front of the user.
If not, then proceed with the next steps:
5) Press the “+” or “-“ (for a faster movement) and “+1”, “-1” (for a slower movement) on the
software for moving the plate clockwise or under clock wise until the position 1 is placed in front
of the catalyst entrance.