EasyManua.ls Logo

Milestone DMA-80 evo - 3 DMA-80 EVO GENERAL INFORMATION; 3.1 Technical data

Milestone DMA-80 evo
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Milestone Srl MA213-003 – DMA-80 EVO – User Manual
34
3 DMA-80 EVO GENERAL INFORMATION
3.1 Technical data
Operating principle
Sample thermal decomposition, amalgamation, atomic absorption with dual beam option
Light source
One low pressure mercury lamp.
Wavelength
253.65 nm.
Detectors
UV enhanced photodiodes,
Instrument controller
Terminal 660
Touch-screen industrial grade controller 6.5” screen with 65.000 colors; VGA resolution 640x480
for sharp process graphic.
5 USB port, 1 LAN connection for network, 2 RS 232 ports for external devices, 2 video port.
Methods and runs saved on PC-compatible USB pen-drive.
Personal Computer (minimum requirements)
CPU Pentium III 800.
RAM 256 MB.
20 MB free space on hard disk.
1 USB port, 1 CD-ROM reader, operating system Windows 7, XP, Vista or 10.
Instrument software
Milestone easyCONTROL
®
LIMS compatible.
Compliant to FDA regulations 21 CFR Part 11.
Detection limit
for solids and liquids
0.003 ng (dual-cell)
0.001 ng (dual-cell double beam)
0.0003 ng (tri-cell double beam)
Typical analysis time
5 minutes: dual cell, tri cell with or without double beam option
8-9 minutes: wide range
Typical precision
Less than 1.5%: dual cell, tri cell with or without double beam option at 10ng
Less than 5%: wide range at 5000ng
Working ranges
for solids and liquids
Automatic switch-over.
Dual-cell: Low range: 0.03 to 20 ng. High range: 20 to 1500 ng.
Wide Range: 10 ng to 20000 ng
Dual-cell Double beam: Low range: 0.01 to 20 ng. High range: 20 to 1500 ng.
Tri-cell Double beam: Low range: 0.003 to 5 ng. Medium range: 0.01 to 20 ng. High range: 20 to
1500 ng.

Table of Contents