Remove the catalyst and amalgamator.
You should remove the two parts separately
because the two silicon joints on the amalgamator
tube will not allow the amalgamator to pass
through the coil.
In come case it is necessary to cut the coil
between amalgamator and catalyst tube.
Install a new catalyst tube with a new o-ring
placed 5mm from the catalyst inlet, and a new
silicon joint placed on the tip.
Wait to connect the amalgamator.
The new catalyst needs to be drained
Once the new tube has been installed, close the
flange, and push in the catalyst oven block.