A5/A3â„¢ Service Manual 046-001141-00 2 - 25
Installation Guide Functional Tests
FIGURE 2-34 Automatic Circuit Leak Test Completed
2. Set up the machine as per the instructions on the screen. Then, select Continue to execute
manual circuit leak test.
3. The ongoing manual circuit leak test is as shown below. You can select Cancel to cancel the
ongoing leak test.
FIGURE 2-35 Manual Circuit Leak Test In Progress
The following screen is displayed if the manual circuit leak test is failed. If so, you must perform the
test again.