1-2
1.2.1 Warnings
WARNING
All installation operations, expansions, changes, modifications and repairs of
this product are conducted by Mindray authorized personnel.
There is high voltage inside the equipment. Never disassemble the
equipment before it is disconnected from the AC power source or the
battery.
When you disassemble/reassemble a parameter module, a patient leakage
current test must be performed before it is used again for monitoring.
The equipment must be connected to a properly installed power outlet with
protective earth contacts only. If the installation does not provide for a
protective earth conductor, disconnect it from the power line and operate it
on battery power, if possible.
Disposal of the packaging material should observe the applicable waste
control regulations and keeping it out of children’s reach.
1.2.2 Cautions
CAUTION
Make sure that no electromagnetic radiation interferes with the
performance of the equipment when preparing to carry out performance
tests. Mobile phone, X-ray equipment or MRI devices are a possible source of
interference as they may emit higher levels of electromagnetic radiation.
Before connecting the receiver to the power line, check that the voltage and
frequency ratings of the power line are the same as those indicated on the
unit’s label or in this manual.
Protect the equipment from damage caused by drop, impact, strong
vibration or other mechanical force during servicing.
1.2.3 Notes
NOTE
Refer to the operator’s manual for detailed operation and other information.
1.3 Equipment Symbols
See theapplicable Operator’s Manual for information about the symbols used on this
product and its packaging.