6-9
3. Unscrew the two M3 screws and remove the power management board assembly.
6.3.7 Removing the Parameter Connector Panel Assembly (With CO
2
,
Only Available for Accutorr 7)
1. Lay the rear housing assembly of the equipment on a table.
2. Pull out the AC receptacle, SpO
2
cable, and then the silicon tube, gas inlet tube, gas
outlet tube and CO
2
connection cable.
2 M3 screws
Power
management
board
Parameter connector panel assembly
SpO
2
signal
cable
Gas inlet tube
CO
2
connection
cable
Gas outlet
tube
Silicon tube