4-7
6. In the [Calibrate CO2] menu, the measured CO2 concentration is displayed. After the
measured CO
2 concentration becomes stable, select [Calibrate CO2] to calibrate the
CO
2 module.
If the calibration is completed successfully, the message [Calibration Completed!] is
displayed in the [Calibrate CO2] menu. If the calibration fails, the message [Calibration
Failed] is displayed. In this case, attempt to calibrate the CO2 module again.
4.2.6 Temperature Test
Required Tool:
Thermostatic oil tank (HART 7102 recommended) or Water Bath and Precision
Thermometer for reference
Test procedure:
1. Set the temperature of the oil tank or water bath to 37°C and conduct the test
after the temperature stabilizes.
2. Set temperature unit to °C.
3. Select temperature parameter area to access [
Temp Setup] menu, and then set
temperature type to [
Monitor].
4. In Monitor Mode, remove the Temperature probe from the probe sheath, insert a
probe cover, and place the probe into the oil tank or water bath.
5. Wait till the Temp value displayed on the monitor stabilizes. Verify that the
displayed value is ±0.1°C of the oil bath thermostat setting or reference
thermometer.
Contact our Technical Support Department if the Temperature test fails.
NOTE
Due to the different application environment and the test object in vivo and
vitro conditions, there are deviations in the measurement result. The
maximum deviation of 2.5Ԩ may exist in predictive Temperature
measurement by liquid bath techniques.

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals