5-10
Symptoms Possible Cause Troubleshooting
stopped" is reported
or some parameters
cannot be used
Parameter module
defective
Replace the corresponding module.
Converter board
defective inside the
module
Replace corresponding converter
board.
"XX" indicates the configured modules, such as NIBP, SpO2 and Temperature.
5.6.13 CO
2
Module Problems (Only Available for Accutorr 7)
5.6.14 Technical Alarm Messages
Please refer to the operator’s manual.
Symptoms Possible Cause Troubleshooting
CO
2
Sensor High
Temp
Ambient temperature
is too high or there is a
module failure.
Lower the operating temperature. If
the alarm persists, the CO
2
module
may have a failure. Change the CO
2
module.
CO
2
Zero Failed The zeroing failed.
Wait for the next successful zeroing,
or repower the module.
CO
2
No Filterline
The mini water trap
disconnected.
Make sure that the mini watertrap
and sampling line is connected.
CO
2
FilterLine
Occluded
The airway or mini
water tarp is occluded.
Check the airway and remove the
occlusion.
CO
2
frequently
zeroes during the
first hour after
start-up.
The CO
2
module
temperature is not
stable.
No action is required. The zeroing
does not affect measurement
accuracy. The zeroing frequency will
decrease over time.
To Purchase, Visit Avobus.com or call 1-800-674-3655

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals