4-6
3. Open the valve to start flowing CO
2 and make sure that there is sufficient flow to vent
to atmosphere.
4. Verify the realtime CO
2 value is within 5.0 ± 0.3% in the [Calibrate CO2] menu.
CO2 Calibration
Tools required:
A steel gas cylinder with 5% CO
2, 21% O2, and balance gas N2 (P/N 0075-00-0033-01)
T-shape connector
Tubing
1. Select [
Main] [Maintenance >>][User Maintenance>>]enter the required
password [
Module Maintenance>>][Calibrate CO2>>].
2. In the [Calibrate CO2] menu, select [Zero].
3. After the zero calibration is finished successfully, connect the equipment as follows:
4. Open the valve to start flowing CO
2 and make sure that there is sufficient flow to vent
to atmosphere.
5. In the [Calibrate CO2] menu, enter the CO
2 concentration in the [CO2] field.
Monitor
Tubing
Cylinder
T
-shape connector
Relief valve
Monitor
Tubing
Cylinder
T-shape connector
Relief valve
To Purchase, Visit Avobus.com or call 1-800-674-3655

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals