Chapter7 Troubleshooting
Solution:
Stop using the recorder. If the error repeats, call Mindray Customer Service
Department or the distributor.
7.2.17 Press Bar Up
Possible causes: The press bar of the recorder is up.
Solution:
Push it back If the error still remains, call Mindray Customer Service
Department or the distributor.
7.2.18 Lyse Out
Possible causes: insufficient lyse or wrong lyse volume setting.
Solution:
Check if there is sufficient lyse left. If so, access Setup
→
Settings
→
Reagents and adjust the remaining lyse volume; if not, change a new container
of lyse.
7.2.19 Diluent Expired
The diluent has expired or its expiration date is not correctly set.
Check whether the diluent has expired. If so, change a new container of diluent;
if not, access Setup
→
Settings
→
Reagents and adjust the expiration date.
7.2.20 Rinse Expired
Possible causes: rinse has expired or its expiration date is not correctly set.
Solution:
Check the rinse. If it is expired, change a new container of rinse; if not, access
Setup
→
Settings
→
Reagents and adjust the expiration date.
7.2.21 Lyse Expired
The lyse has expired or its expiration date is not correctly set.
Check whether the lyse has expired. If so, change a new container of lyse; if not,
access Setup
→
Settings
→
Reagents and adjust the expiration date.
7.2.22 Filter Error
Something is wrong with the filter of the vacuum chamber.
Solution:
Access Service
→
Self-test
→
Tubing and test the filter. The error is cleared
if the result is normal. If not, call Mindray Customer Service Department or the
distributor.
7.2.23 Real-Time Clock Error
Something is wrong with the clock.
Access Setup
→
Settings
→
Date & Time and set the time and date and
restart the analyzer to enable the new settings. If the error still remains, call
BC-2800 Auto Hematology Analyzer (V1.0) 7-7