Histograms and Pulse Graphs
Figure 4-5
6. A histogram that does not show differential result because of severe
interference (check the pulse graph to determine whether there is
interference) in the WBC channel.
Figure 4-6
7. No lyse reagent or poor hemolysis
Figure 4-7
8. Increased granulocytes
Figure 4-8
9. Increased lymphocytes
Figure 4-9
10. Tumor patient
Figure 4-10
11. Increased monocytes
Figure 4-11
4-2 BC-2800 Auto Hematology Analyzer (V1.0)