Disassembling the Analyzer
1. Follow steps 1~6 according to Section 3.3.1.
2. Pull the loosen sample probe upward until it out of the wipe block, then remove
the wipe block and disconnect the tubings from the it (pay attention to the
correspondence between the tubings and the connectors), as the figure below
shows. Install a new wipe block and connect the tubings (the tubing marked with
a black line should connect to the bottom connector). When the connection is
done, place the wipe block back to its original position.
3. Follow steps 7~9 according to Section 3.3.1 to fix the sample probe.
Figure3-25 Installing probe wipe
BC-2800 Auto Hematology Analyzer (V1.0) 3- 15