Chapter 6 Maintenance
desired screen from there.
6.2.1 Diluent Prime
Biohazard
Wear standard laboratory attire (including rubber gloves) and
follow safe laboratory procedures when handling any material in
the laboratory.
You should perform the Diluent prime procedure to prime the diluent tubing
with diluent when
there are bubbles in the tubing; or
the diluent in the tubing is contaminated; or
the old diluent ran out and a new container of diluent is installed.
Follow the steps given below to do so
At the Maintenance screen, press the appropriate arrow keys ([↑][↓][←][→])
to move the cursor to Diluent Prime.
Press [ENTER] to prime the tubing with diluent and the priming progress will be
displayed at the bottom of the screen, as Figure 6-3 shows.
When the priming is done, the screen will return to the initial state.
Figure 6-3 Diluent priming screen
BC-2800 Auto Hematology Analyzer (V1.0) 6-5