Disassembling the Analyzer
Figure3-9
3.2.7 Removing the power shielding box
As the figure below shows, remove the retaining screws (totally 3) to remove the
box.
Figure3-10
3.2.8Removing the driving board, analog board, CPU board and
power board
As the figure below shows, remove the retaining screws (totally 18) to remove
the boards.