Chapter7 Troubleshooting
7.2.30 Diluent Out
Possible causes: insufficient diluent or wrong diluent volume setting.
Solution:
Check if there is sufficient diluent left. If so, access Setup
→
Settings
→
Reagents and adjust the remaining diluent volume; if not, change a new
container of diluent.
7.2.31 Rinse Out
Possible causes: insufficient rinse or wrong rinse volume setting.
Solution:
Check if there is sufficient rinse left. If so, access Setup
→
Settings
→
Reagents and adjust the remaining rinse volume; if not, change a new container
of rinse.
7.2.32 Waste Full
The waste container is full.
Empty the container, or change a new container to receive the waste and re-set
the waste container volume.
7.2.33 File Error
Something is wrong with file saving.
Turn off the analyzer and call Mindray Customer Service Department or the
distributor.
7.2.34 Dynamic Memory Error
Something is wrong with the system memory.
Turn off the analyzer and call Mindray Customer Service Department or the
distributor.
BC-2800 Auto Hematology Analyzer (V1.0) 7-9