Disassembling the Analyzer
3.2 Disassembling the Main Unit
3.2.1 Removing the left, right and top covers
As the figure below shows, remove the retaining screws (two for each cover) with
a Philips screwdriver and pull the covers away.
Figure3-4
Removing the left, right and top covers
3.2.2 Removing the back plate and the fixing plate of the power
module:
As the figure below shows, remove the retaining screws (totally 10) to remove
the plates.
BC-2800 Auto Hematology Analyzer (V1.0) 3- 3