Chapter 6 Maintenance
6.2.2 Rinse Prime
Biohazard
Wear standard laboratory attire (including rubber gloves) and
follow safe laboratory procedures when handling any material in
the laboratory.
You should perform the Rinse prime procedure to prime the rinse tubing with
rinse when
there are bubbles in the tubing; or
the rinse in the tubing is contaminated; or
the old rise ran out and a new container of rinse is installed.
Follow the steps given below to do so:
At the Maintenance screen, press the appropriate arrow keys ([][][][])
to move the cursor to Rinse prime.
Press [ENTER] to prime the tubing with rinse and the priming progress will be
displayed at the bottom of the screen, as Figure 6-4 shows.
When the priming is done, the screen will return to the initial state.
Figure 6-4 Rinse Priming screen
BC-2800 Auto Hematology Analyzer (V1.0) 6-6