Chapter 6 Maintenance
When the soaking is done, press [ENTER] to flush the bath and tubing, after
which screen will return to the initial state.
To make sure this analyzer functions normally, every time the accumulated
analyzed samples reach 200, a dialog box will pop up to remind you to perform
the Probe cleanser cleaning procedure, as Figure 6-8 shows. If you want to do
so, move the cursor to Yes and press [ENTER]; otherwise, move the cursor to
No and press [ENTER].
Figure 6-8 Recommending probe cleanser cleaning
6.2.7 E-Z Cleanser Cleaning
Biohazard
Wear standard laboratory attire (including rubber gloves) and
follow safe laboratory procedures when handling any material in
the laboratory.
You can use the E-Z cleanser, an enzyme based, isotonic cleaning solution and
wetting agent, to clean the tubing and bath by performing the E-Z cleanser
cleaning procedure.
Follow the steps given below to perform the procedure:
At the Maintenance screen, press the appropriate arrow keys ([↑][↓][←][→])
to move the cursor to E-Z cleanser cleaning.
Present the cleanser to the probe and press [ENTER] to aspirate the cleanser.
Remove the cleanser after the probe has risen up. This analyzer will
automatically soak the bath and tubing with the aspirated cleanser.
BC-2800 Auto Hematology Analyzer (V1.0) 6-10