Chapter 6 Maintenance
Figure 6-7 Flushing aperture
6.2.6 Probe Cleanser Cleaning
Biohazard
Wear standard laboratory attire (including rubber gloves) and
follow safe laboratory procedures when handling any material in
the laboratory.
You can soak the bath and tubing with the probe cleanser, an alkaline detergent,
by performing the Probe cleanser cleaning procedure. If your analyzer is to
run 24 hours a day, you should perform this procedure every 3 days. If you
follow the shutdown procedure to turn off your analyzer everyday, you should
perform this procedure every week.
Follow the steps given below to do so:
At the Maintenance screen, press the appropriate arrow keys ([][][][])
to move the cursor to Probe cleanser cleaning.
Present the cleanser to the probe and press [ENTER] to aspirate the cleanser.
Remove the cleanser after the probe has risen up. This analyzer will
automatically soak the bath and tubing with the aspirated cleanser.
The soaking process will last about 5 minutes and you may press [ENTER] to
stop the process any time. Note that a shortened soaking process may not be as
effective as a complete one.
BC-2800 Auto Hematology Analyzer (V1.0) 6-9