Disassembling the Analyzer
Figure
3-23 Insert localizer
5. Insert the probe into the wipe block until it reaches the localizer, as the figure
below shows.
Figure3-24 Inserting sample probe into wipe block
6. Re-tighten the screw to fix the probe and remove the localizer.
3.2.17 Replacing Wipe block
Biohazard
Wear standard laboratory attire (including rubber gloves) and
follow safe laboratory procedures when handling any material in
the laboratory.
Warning
The probe is sharp and may contain biohazard materials, including
controls and calibrators. Avoid any unnecessary contact with the
probe and probe area.
Follow the steps given below to replace the wipe block:
3-14 BC-2800 Auto Hematology Analyzer (V1.0)