Chapter 6 Maintenance
At the Maintenance screen, press the appropriate arrow keys ([][][][])
to move the cursor to Drain bath.
Press [ENTER] to start the procedure and the screen will display the progress
bar..
When the bath is drained, the screen displays Drain bath, as Figure 6-13
shows..
Check the bath and its lower hose for residual liquid. If there is, turn off the
analyzer and call the Mindray Customer Service Department or the distributor for
repair; if not, press [ENTER] and the analyzer will prime the bath with diluent.
When the priming is done, the screen will return to the initial state and you
should try to find other reasons the contribute to the abnormal results.
Figure6-13 Drain Bath
6.2.11 Draining Tubing
If this analyzer is not to be used for a long time or it is to be maintained, be sure
to perform the Drain tubing procedure.
Follow the steps given below to do so
At the Maintenance screen, press the appropriate arrow keys
([][][][])to move the cursor to Drain tubing.
Press [ENTER] and follow the displayed instructions to remove all the tubes,
except for the one for the waste, from this analyzer.
Press [ENTER] to start the draining procedure.
When the draining is done, the screen will display Turn off this analyzer and
you should turn off this analyzer as instructed.
BC-2800 Auto Hematology Analyzer (V1.0) 6-14