Chapter 6 Maintenance
At the Maintenance screen, press the appropriate arrow keys ([][][][])
to move the cursor to Zap aperture.
Press [ENTER] to zap the aperture and the zapping progress will be displayed at
the bottom of the screen, as Figure 6-6 shows.
When the zapping is done, the screen will return to the initial state.
Figure 6-6 Zapping aperture
6.2.5 Flush Aperture
You can perform the Flush aperture procedure to assist zapping the aperture.
Follow the steps given below to perform the procedure
At the Maintenance screen, press the appropriate arrow keys ([][][][])
to move the cursor to Flush aperture.
Press [ENTER] to flush the aperture and the flushing progress will be displayed
at the bottom of the screen, as Figure 6-7 shows.
When the flushing is done, the screen will return to the initial state.
BC-2800 Auto Hematology Analyzer (V1.0) 6-8