Histograms and Pulse Graphs
Figure 4-36 Insufficient liquid in RBC bath
during PLT counting
Figure 4-38 Interference in WBC channel
from tubing. Feature: data length
increases, the base line of signal is not
stable.
Figure 4-37 Interference in PLT channel
from tubing. Feature: data length
increases, the base line of signal is not
stable.
Figure 4-39 Interference in RBC channel
from tubing. Feature: data length
increases, the base line of signal is not
stable.
4-8 BC-2800 Auto Hematology Analyzer (V1.0)