EMC
B - 4 BP10 Operator’s Manual
Guidance and Declaration — Electromagnetic Immunity
The system is intended for use in the electromagnetic environment specified below.
The user of the system should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic Environment
- Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
Transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Not applicable Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC
61000-4-5
± 1 kV lines(s) to
line(s)
± 2 kV line(s) to earth
Not applicable
Voltage dips,
Short
interruptions
and voltage
variation on
power supply
input lines IEC
61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 seconds
Not applicable Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of our product requires
continued operation during
power mains interruptions, it is
recommended that our product
be powered from an
uninterruptible power supply or
a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Note: U
T
is the A.C. mains voltage prior to application of the test level.