34
No. 99MBC139A
3 Operation Procedure
■Advanced parameters
The advanced parameters increase the scope of utility of the product. These parameters make it possi-
ble to make settings such as selection of the tolerance judgment function and I/O-related settings.
No. Set value
Per-
axis
setting
Set value: Operation
Default
value
Description
01 Key protect
N/A
0: No key protect
1: Key protect
0
Operation of keys other than those
used for setting parameters can be
disabled to prevent operation errors.
Tips
Operation by external input cannot be
disabled.
02
Origin initial-
ization
(Origin clear)
N/A
0: Do not initialize
1: Initialize
0
When a Linear Gage with origin
mark is connected, the origin can
be initialized without cycling on the
power.
03
Display mode
selection
N/A
Set
value
Ch. 1 Ch. 2
0
A-axis
counter
B-axis
counter
1
Sum
(A+B)
B-axis
counter
2
Dier-
ence
(A-B)
B-axis
counter
3
A-axis
counter
Sum
(A+ B)
4
A-axis
counter
Dier-
ence
(A-B)
5
A-axis
speed
B-axis
speed
6
A-axis
counter
A-axis
speed
7
B-axis
counter
B-axis
speed
0
Selects the values to be displayed
on Ch. 1 and Ch. 2.
Tips
• When using only the A-axis or the
B-axis, set 6 or 7.
• If the displayed content is the same
both before and after changing the
setting, the tolerance value, preset
value and preset state settings are
maintained.
• In order to set the speed display,
set the speed sampling cycle with
Parameter Number 17.
• The speed display unit is mm/s (or
in/s). The display of the lower 1 or
2digitsmaybexeddependingon
the speed sampling cycle.
• When the minimum value (MIN) is
selected for Peak mode, the speed
display indicates the maximum
speed in the reverse direction.
• Because speed display is simpli-
ed,itisnotsuitableforfeedback
control.