34
No. 99MBC139A
3 Operation Procedure
■Advanced parameters
The advanced parameters increase the scope of utility of the product. These parameters make it possi-
ble to make settings such as selection of the tolerance judgment function and I/O-related settings.
No. Set value
Per-
axis
setting
Set value: Operation
Default
value
Description
01 Key protect
N/A
0: No key protect
1: Key protect
0
Operation of keys other than those
used for setting parameters can be
disabled to prevent operation errors.
Tips
Operation by external input cannot be
disabled.
02
Origin initial-
ization
(Origin clear)
N/A
0: Do not initialize
1: Initialize
0
When a Linear Gage with origin
mark is connected, the origin can
be initialized without cycling on the
power.
03
Display mode
selection
N/A
Set
value
Ch. 1 Ch. 2
0
A-axis
counter
B-axis
counter
1
Sum
(A+B)
B-axis
counter
2
Di󰀨er-
ence
(A-B)
B-axis
counter
3
A-axis
counter
Sum
(A+ B)
4
A-axis
counter
Di󰀨er-
ence
(A-B)
5
A-axis
speed
B-axis
speed
6
A-axis
counter
A-axis
speed
7
B-axis
counter
B-axis
speed
0
Selects the values to be displayed
on Ch. 1 and Ch. 2.
Tips
When using only the A-axis or the
B-axis, set 6 or 7.
If the displayed content is the same
both before and after changing the
setting, the tolerance value, preset
value and preset state settings are
maintained.
In order to set the speed display,
set the speed sampling cycle with
Parameter Number 17.
The speed display unit is mm/s (or
in/s). The display of the lower 1 or
2digitsmaybexeddependingon
the speed sampling cycle.
When the minimum value (MIN) is
selected for Peak mode, the speed
display indicates the maximum
speed in the reverse direction.
Because speed display is simpli-
ed,itisnotsuitableforfeedback
control.