43 Amplification error Overcurrent flowed into the motor
driver.
Restart the analysis software and
controller.
44 System error Disconnect between the main unit
and the controller
Turn the power off and check for
correct wiring connections between
the main unit and the controller.
5* USB
communication
error *)
Communication error between the
PC and the controller
Check that the USB cable is correctly
connected between the PC and the
controller.
6* Remote
communication
error
Communication failure between
the remote box and the controller
Check that the remote box and the
controller are correctly connected.
No. 99MBB187A
5-6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com