OPR360US H0711
6-3
Cleaning and Maintenance
6
Exhalation Manifold
Exhalation Flow Sensor
The following diagram and instructions are for the removal and
reinstallation of the exhalation flow sensor.
1.
Open the front panel door on the lower left front of the ventilator to
expose exhalation valve and flow sensor. See Figure 6-2
2. With a gentle twisting motion, pull the plastic flow sensor away
from the outlet of the exhalation valve. See Figure 6-3.
Figure 6-1 Fan Filter
Figure 6-2 Exhalation Valve and Flow Sensor